Journal of Zhejiang University SCIENCE B
ISSN 1673-1581(Print), 1862-1783(Online), Monthly
2005 Vol.6 No.12 P.1148-1152
Determination of theophylline concentration in serum by chemiluminescent immunoassay
Abstract: Objective: This study aimed to establish chemiluminescent immunoassay (CLIA) for quantitative determination of theophylline levels in human serum. Methods: To measure the concentration of theophylline (n=122) and evaluate the assay. Results: The linear range of the CLIA method was 0.51~40 mg/L (Y=1.02X+0.44, r=0.995). The intra and inter CV (coefficient variance) of CLIA were 3.20% and 3.57%, respectively. The average recovery rate was 102.3%. This method was free from interference by brilirubin (<200 μmol/L), hemoglobin (<10 g/L), and triglycerides (<15 mmol/L). Conclusion: This method is simple, convenient and precise for clinical pharmacokinetics study of theophylline.
Key words: Theophylline, Chemiluminescent immunoassay (CLIA), Fluorescence polarization immunoassay (FPIA)
References:
Open peer comments: Debate/Discuss/Question/Opinion
<1>
DOI:
10.1631/jzus.2005.B1148
CLC number:
R331.1
Download Full Text:
Downloaded:
2799
Clicked:
5787
Cited:
7
On-line Access:
2024-08-27
Received:
2023-10-17
Revision Accepted:
2024-05-08
Crosschecked: