Simultaneous detection of seven phenolic acids in Danshen injection using HPLC with ultraviolet detector
Abstract: A high-performance liquid chromatographic (HPLC) method with ultraviolet (UV) detector had been developed for simultaneous quantification of danshensu, protocatechuic aldehyde, caffeic acid, salvianolic acid D, rosmarinic acid, salvianolic acid B and salvianolic acid A in Danshen injection. According to the UV spectra of these components, three detection wavelengths have been selected as follows: 280 nm for danshensu and protocatechuic aldehyde, 326 nm for caffeic acid, salvianolic acid D and rosmarinic acid, 286 nm for salvianolic acid B and salvianolic acid A. The limit of detection (LOD) was improved to be in the range of 0.008~0.160 μg/ml. Moreover, excellent linear behavior over the investigated concentration range was observed, with R>0.999 for all the analytes.
Key words: High-performance liquid chromatography (HPLC), Danshen injection (DSI), Phenolic acids, Multi-wavelengths, Quantification